Description
Trilogy 100 makes invasive and noninvasive home ventilation less complicated with a simplified user interface and Respironics’ proven BiPAP technology for greater versatility. Its lightweight design and internal and detachable batteries make portability easier for patients on the go.
Light, versatile and easy-to-use, that’s the Trilogy100 portable ventilator at its essence. With its light weight and proven technology, Trilogy100 makes invasive and non-invasive treatment less complicated for a wide range of adult and pediatric patients.
Connected
Keeping care teams in the know
Trilogy is now connected to Care Orchestrator, our HIPAA-compliant cloud-based patient management system. With the ability to transmit data wirelessly and remotely, Connected Trilogy helps providers deliver quality respiratory care, gain efficiencies, potentially avoid unnecessary home visits, improve care coordination, streamline workflow and prioritize patient need.Reliable and portable
Continuous respiratory support for added freedom and mobility
Trilogy provides respiratory support to meet your prescribed needs, up to 24 hours a day. Trilogy is portable with internal, hot-swappable detachable, and external batteries to provide greater freedom and mobility with continuous ventilation. The easy-to-use carry bag is wheelchair-mountable and designed so the screen is always visible.Flexible
Easy to use and adaptable for your patients’ changing needs
Trilogy is designed with easy-to-read and navigate screens, multiple Rx settings and on-screen waveforms to help optimize treatment. Trilogy’s pressure- and volume-control features allow you to provide more types of therapy with one device. The ability to use multiple modes for patients who require varying levels of ventilation, allows greater flexibility in care.Volume targeting
Adapting with every breath
AVAPS-AE–an auto-titration mode of non-invasive ventilation with multiple algorithms that work simultaneously–meets the changing needs of your patients. AVAPS automatically adjusts to breathing, while Auto EPAP proactively adjusts to the lowest effective pressure for peak comfort. Auto Backup delays a machine breath until the patient has exhaled to reduce air trapping.Mouthpiece ventilation
Delivering on-demand ventilation
With no inspiratory effort required to trigger a breath, our unique ‘kiss’ trigger detects when a patient engages and disengages from the mouthpiece to deliver on-demand ventilation with the security of patient alarms.Digital Auto-Trak
When sensitivity is needed most
This adaptive algorithm automatically adjusts the triggering and cycling of the ventilator, while compensating for leaks. Digital Auto-Trak’s capability contributes to the patient’s synchrony and comfort, without requiring manual adjustments. The Sensitive Auto-Trak setting provides an enhanced triggering response for patients with minimal respiratory effort.
| SPECIFICATIONS |
| Environmental | |
| Temperature | 5°C – 40°C |
| Relative humidity | 15% – 95% |
| Atmospheric pressure | 110 kPa to 60 kPa |
| Physical | |
| Weight | 11 lbs, 5 kgs |
| Size | 6.6″L x 11.2″W x 9.3″H (16.68 cm L x 28.45 cm W x 23.52 cm H) |
| Controls | |
| Circuit types | Active with PAP, passive |
| Ventilation types | Volume control, pressure control, bilevel |
| Volume modes | AC, SIMV (w/PS), CV |
| Pressure modes | CPAP, S, S/T, T, PC-SIMV (w/PS) |
| IPAP | 4 – 50 cm H20 |
| EPAP | 0 – 25 cm H20 for active circuits 4 – 25 cm H20 for passive circuits |
| CPAP | 4 – 20 cm H20 for passive circuits |
| PEEP | 0 – 25 cm H20 for active circuits 4 – 25 cm H20 for passive circuits |
| Pressure support differential | 0 – 30 cm H20 |
| Measured patient parameters | |
| Tidal volume | 50 – 2000 ml |
| Breath rate | 0 – 60 for AC mode 1 – 60 for all other modes |
| Inspiratory time | 0.3 – 5.0 secs |
| Rise time | 1 – 6 |
| Ramp start pressure | 0 – 25 cm H20 for active circuits 4 – 25 cm H20 for passive circuits 4 – 19 cm H20 in CPAP mode |
| Ramp length | Off, 5 – 45 minutes |
| Flex | Off, 1 – 3 |
| Flow trigger sensitivity | 1 – 9 l/min |
| Flow cycle | 10% – 40% |
| Apnea rate | 4 – 60 BPM |
| Tidal volume | 0 – 2000 ml |
| Minute ventilation | 0 – 99 l/min |
| Leak rate | 0 – 200 l/min |
| Respiratory rate | 0 – 80 BPM |
| Peak inspiratory flow | 0 – 200 l/min |
| Peak inspiratory pressure | 0 – 99 cm H20 |
| Mean airway pressure | 0 – 99 cm H20 |
| Percentage patient triggered breaths | 0 – 100% |
| I:E ratio | 9.9:1 to 1:9.9 |
| Alarms | |
| Circuit disconnect | Off, 10 – 60 secs |
| Apnea | Off, 10 – 60 secs |
| High tidal volume | Off, 50 – 2000 ml |
| Low tidal volume | Off, 50 – 2000 ml |
| High minute ventilation | Off, 1 – 99 l/min |
| Low minute ventilation | Off, 1 – 99 l/min |
| High respiratory rate | Off, 4 – 80 BPM |
| Low respiratory rate | Off, 4 – 80 BPM |
INCLUDES:Rechargeable Battery, andpower supply.






Reviews